A PTFE capillary flow acceptance requirement should state the measured flow and its allowed range under one defined fluid, temperature, differential pressure, specimen length, conditioning state, flow direction and measurement method. Nominal inner diameter remains necessary for manufacture and inspection, but it does not by itself define the delivered hydraulic restriction.
This guide is an original specification and reporting framework. It is not Pengrowth flow-test data, a standard test method or a promise that a particular test service is available. For dimensional sourcing, start with the PTFE capillary tubing product page and the existing ID, OD and tolerance guide.
Why inner diameter alone does not close a flow requirement
Flow through a small bore is a system response. Bore size and length matter, but so do the fluid’s viscosity, actual temperature, pressure boundary, bore variation, roundness, cut ends, connectors, wetting condition and measurement chain. A drawing that lists only “ID 0.50 mm” leaves the buyer and supplier free to test different physical questions.
Dimensional acceptance and functional flow acceptance answer different questions. A dimensional result describes the inspected locations and method. A flow result integrates the hydraulic behaviour of the tested specimen and fixture under stated conditions. Use both where the assembly needs both fit and restriction; do not silently replace one with the other.
Define the unit of supply as well. A cut piece, a coil segment and a finished assembly may contain different effective lengths and end conditions. State where length begins and ends, whether connectors are included, and whether the result applies to each piece, a sample from a lot or another declared sampling plan.
Define the measurand and decision before building the fixture
Choose mass flow or volumetric flow and state the reporting units. If mass collected over time is converted to volume, identify the fluid density and the temperature or other condition to which it applies. The official description of ASME MFC-9M illustrates the mass-over-time principle and the need for density in a volumetric conversion; it does not make every bench collection setup compliant with that standard.
Write the acceptance band, rounding rule and uncertainty decision rule before seeing results. For example, the contract may require the reported value plus its stated uncertainty to be considered against agreed limits. The exact rule belongs to the parties and the risk of the application. Do not decide after testing whether uncertainty is ignored, subtracted or used to widen a tolerance.
ISO 5168:2005 establishes general principles for evaluating uncertainty in fluid-flow measurements. Its public page does not supply a complete low-flow test procedure. The laboratory or responsible engineer must build an uncertainty budget for the chosen range and instruments rather than copy a generic accuracy figure.
Lock the fluid, temperature, pressure and specimen boundary
| Item | Define before testing | Why it changes the meaning |
|---|---|---|
| Fluid | Identity, grade/composition, preparation and reuse rule | Viscosity, density, contamination and gas content can change the result |
| Temperature | Target, tolerance, measurement location and stabilization rule | Fluid properties and the test system can be temperature-sensitive |
| Pressure boundary | Upstream and downstream measurement points, target differential and tolerance | Supply setting is not necessarily pressure drop across the specimen |
| Specimen | Effective length, end preparation, connectors and installed geometry | Length and local restrictions affect the measured response |
| State | Conditioning, wetting/priming, bubble-removal and direction rule | Start-up history can affect stability and repeatability |
| Decision | Sampling, repeats, acceptance limits and uncertainty rule | A number without a predeclared decision rule cannot close acceptance |
Measure pressure at points that bracket the intended specimen boundary. A regulator set point upstream of valves and fittings does not establish the pressure drop across the capillary. Characterize a fixture blank or justified bypass so tubing resistance is not confused with hoses, filters, connectors or the measuring device. Record the blank configuration and how its contribution is handled.
The NIST description of its liquid flow standards notes both traceable calibration work and research into temperature sensitivity. NIST’s piston-prover publication documents working-fluid properties, distributed temperature measurements, corrections and uncertainty analysis. Those facilities operate in different ranges; the transferable lesson is to document the measurement system and actual state, not to claim equivalence to a NIST calibration.
Control conditioning, wetting, bubbles and direction
Write a conditioning and wetting sequence that the actual fluid system can reproduce. Identify whether specimens are tested as received, dried, equilibrated in a stated environment or preconditioned with the test fluid. State the priming volume or time and the criterion used to decide that visible or detected gas has been removed. “Fully wetted” is not a procedure.
Define a stabilization rule without inventing a universal waiting time. A rule might evaluate the change in consecutive readings over a stated interval, subject to the resolution and response time of the instruments. Preserve the start-up record where it matters. If the signal never meets the rule, report that outcome rather than selecting a convenient later value.
Mark the specimen direction and either test one nominated direction or run a predeclared reversal check. A direction difference may point to end preparation, debris, an asymmetric connector or bore variation; it does not by itself prove the cause. Do not average opposite directions until the protocol says that they represent the same measurand.
Choose a measurement chain that resolves the acceptance band
A calibrated flow meter, timed gravimetric collection, piston displacement or another justified method may be suitable in a particular range. Selection depends on fluid compatibility, expected flow, response time, evaporation, density knowledge, collection duration and the needed uncertainty. Confirm calibration status over the working range; a full-scale accuracy statement can be misleading near the bottom of a range.
Record instrument identity, calibration reference and date, resolution, acquisition interval and processing. Include pressure, temperature, timing, balance or volume-reference contributions as applicable. Leakage, evaporation, pulsation, trapped gas, drift and zero offset may need experiments or controls rather than a sentence in the report.
Retain individual specimen and repeat values. Report exclusions and equipment alarms. When the measurement uncertainty or observed repeatability is too large relative to the acceptance band, the defensible conclusion is that the setup cannot decide acceptance at that band—not that the specimen passed because the display fell inside it once.
Use the ideal capillary equation as a sensitivity model only
For steady, fully developed laminar flow of an incompressible Newtonian liquid through a straight, rigid, circular tube with no wall slip, the ideal relation can be written:
Q = π × ΔP × d⁴ / (128 × μ × L)
Here Q is volumetric flow rate, ΔP is pressure drop across the modeled tube, d is internal diameter, μ is dynamic viscosity and L is tube length, in a consistent unit system. The fourth power on d explains why a small bore difference can have a much larger modeled effect on flow than the same relative length difference.
The equation is not an acceptance result. The NBS capillary-viscometer monograph discusses Poiseuille relations and corrections associated with real capillary measurements. Entrance and exit effects, noncircular or varying bore, end damage, fixture losses, deformation, slip, non-Newtonian behaviour and an invalid laminar assumption can break the simple model. Gas flow also needs compressibility treatment and is outside this liquid-flow worksheet.
Use the model to ask sensitivity questions and to choose an instrument range. Use controlled measurements to accept product. Do not insert nominal drawing values into the formula and label the output “tested flow,” or use it to promise a production tolerance without supporting capability data.
Complete one blank record for every defined acceptance group
The following is an original, empty planning record. It contains no Pengrowth result, default limit or mandatory test setting.
| Field | Agreed specification | Recorded result / evidence |
|---|---|---|
| Part, lot, sample plan and specimen IDs | ________ | ________ |
| Effective length, ends, connectors and direction | ________ | ________ |
| Fluid identity, preparation, density/viscosity reference | ________ | ________ |
| Temperature target, tolerance, location and actual values | ________ | ________ |
| Upstream/downstream points and differential pressure | ________ | ________ |
| Conditioning, priming, bubble and stabilization rules | ________ | ________ |
| Flow method, instruments, calibration and fixture blank | ________ | ________ |
| Individual repeats, uncertainty and exclusions | ________ | ________ |
| Acceptance band, decision rule and disposition | ________ | ________ |
Send a bounded requirement for manufacturability review
Send the nominal ID and OD, dimensional limits, effective test length, cut/end condition, fluid, temperature, pressure boundary, expected flow range, unit-of-supply definition, sample plan and proposed acceptance rule. The custom PTFE tubing specification guide covers the wider drawing and order inputs.
Ask the supplier to confirm which dimensional characteristics can be controlled and what evidence can accompany the material. If functional flow evidence is required, agree who owns the fixture, method, calibration and disposition before ordering. A request for review does not establish that Pengrowth offers an in-house flow test or that any flow tolerance is achievable.