For a circular, uniform tube bore, geometric volume is V = π × ID² × L / 4. Use internal diameter, not outside diameter. The result describes the tube cavity only: valve passages, fittings and the selected flow path must be considered separately before defining an installed sample loop.
Calculate tube-only volume or length · Understand the limits · Prepare a tubing enquiry
Tube-only geometric calculator
Use consistent units. This tool does not calculate pressure capacity, flow, calibrated injection volume or required flushing.
Use the bore diameter; no Pengrowth size or tolerance is implied.
Straightened tube centreline length, not end-to-end loop span.
Interactive calculation requires JavaScript. You can still calculate manually using the formula below.
Check the units before using the result
When ID and length are in millimetres, the formula returns mm³, numerically equal to µL. A length entered in metres must first be multiplied by 1,000. To calculate tube-only length, use L[mm] = 4 × V[µL] / (π × ID[mm]²).
Hypothetical example: a 0.50 mm circular bore and 1,000 mm tube length give approximately 196.35 µL. A 100 µL tube-only target at the same bore gives approximately 509.30 mm. These decimal places show arithmetic, not a promised cutting precision or volumetric accuracy.
Why ID tolerance matters more than a diameter-only comparison suggests
Volume scales with the square of ID and linearly with length. For an illustrative bore range of 0.48–0.52 mm and length range of 995–1,005 mm, the extreme combinations give about 180.05–213.43 µL. These are hypothetical worst-case bounds, not a statistical confidence interval or an available tolerance. Define drawing limits and the measurement method before ordering.
Geometric volume is not calibrated sample volume
Separate the tube cavity from the complete flow path. Identify the relevant valve position, swept passages and any spaces without through-flow. Do not label every fitting’s internal volume as a deadleg. An injection system can include valve-passage volume in its delivered sample; this depends on the actual assembly and method, not on the tube formula alone. See the instrument-specific context in the IDEX injection-valve instructions.
The model assumes a circular, uniform, undeformed bore. It does not account for bore variation, bubbles, fittings, pressure/temperature deformation or measurement uncertainty. When actual capacity matters, agree how the installed assembly will be measured and accepted.
How much flushing is enough?
There is no universal flushing multiple derived from tube volume. Specify the fluid change, selected valve path, temperature, actual flow and acceptable residual level. Assess measured output against a predefined criterion with an appropriate blank and detection limit. “Not detected” is not the same as zero residue. Swagelok’s representative-sampling guide explains why mixing volumes and deadlegs behave differently.
What to send for a PTFE capillary tubing review
Share the target volume and whether it means tube-only geometry or installed assembly capacity, ID/OD constraints, allowed length range, fluid and service conditions, valve/fitting interfaces and quantity. State any required measurement or acceptance method; do not assume the tubing is supplied as a calibrated injection loop.